Calibration
CALIBRATION · ISO/IEC 17025 No. 0383
Our calibration laboratory is accredited to ISO/IEC 17025 by NSC-TISI (accreditation No. 0383). Every certificate is traceable to the national standard and recognised worldwide through the ilac-MRA.
ilac-MRAInternational mutual recognition arrangement
NAC ThailandNSC-TISI-TIS 17025 · Calibration 0383- ISO/IEC17025Testing & calibration laboratoriesISO/IEC 17025The standard our laboratory is accredited to
Calibration you can trace
TRACEABLE CALIBRATIONNational standard
NIMT Thailand
Reference standard
held and re-certified in our lab
Working standard
what our engineers carry to your site
Your instrument
with certificate
Every certificate we issue traces back to the national standard.
The laboratory
ห้องปฏิบัติการสอบเทียบThe laboratory of CG Engineering calibrates measuring and testing instruments for rubber and plastics, on site at your plant, to ISO/IEC 17025. It is staffed by trained personnel and works to standardised methods with standard equipment traceable to SI units, so results hold up metrologically.
The laboratory was first accredited to ISO/IEC 17025:1999 E (TIS 17025-2000) in 2004, under calibration certification number 0108. The certification was then renewed and updated to ISO/IEC 17025:2005 E on 9 March 2006.
The scope of certification of calibration methods gradually expanded. In 2016 certification number 0108 was withdrawn on relocation of the laboratory, and the lab was recertified to ISO/IEC 17025:2005 on 29 January 2016 by the Thai Industrial Standards Institute (TISI), under calibration certification number 0267.
The laboratory renewed the certificate to ISO/IEC 17025:2017 E (TIS 17025-2018) on 10 January 2019, and improves the quality of its calibration work continuously to serve a growing customer base — from around 15 customers at the beginning to 289 today.
Where we work
- Domestic
- Rubber industry · Plastic industry · Government organisations
- International
- Vietnam · Singapore · Indonesia · Malaysia · Cambodia · Myanmar · China · Ivory Coast · France · England
Calibration scope
ขอบข่ายการสอบเทียบOur laboratory’s calibration scope, by instrument type. Every figure is transcribed from the published scope document — nothing rounded, nothing inferred.
Rheometer (MDR / RPA / ODR)
Rheometer (MDR & RPA)
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 100 to 210 °C | ± 0.50 °C |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-07 by direct comparison to standard Thermometer.
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 0.5 Degree. | ± 1.60% |
| 1.0 Degree | ± 1.60% |
| 3.0 Degree | ± 1.60% |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-15 Verification by Standard Torque
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 100 cpm | ± 0.1 cpm |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-17 by measurement with Standard Stopwatch.
Calibration and measurement capability is expressed as an uncertainty (±) which, for k = 2, provides a level of confidence of approximately 95%.
Rheometer (Only ODR type)
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 100 to 200 °C | ± 0.40 °C |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-06 by direct comparison to standard Thermometer.
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 0.5 Degree. | ± 1.60% |
| 1.0 Degree | ± 1.60% |
| 3.0 Degree | ± 1.60% |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-15 Verification by Standard Torque
Calibration and measurement capability is expressed as an uncertainty (±) which, for k = 2, provides a level of confidence of approximately 95%.
Mooney Viscometer
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 85 to 145 °C | ± 0.35 °C |
| >145 to 160 °C | ± 0.37 °C |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-06 by direct comparison to standard Thermometer.
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| ML1+4@100 °C | ± 0.5 MU |
| ML1+8@100 °C | ± 0.5 MU |
| ML1+4@125 °C | ± 0.5 MU |
| ML1+8@125 °C | ± 0.5 MU |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-16 Verification by Standard material (IRM-241)
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 4 to 15 kg | ± 0.3 g. |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-01 by comparison with standard weight and Electronic balance.
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 2 rpm | ± 0.03 rpm |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-17 by measurement with Standard Stopwatch.
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 1 min | ± 0.1 Seconds |
| 4 min | ± 0.1 Seconds |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-17 by measurement with Standard Stopwatch.
Calibration and measurement capability is expressed as an uncertainty (±) which, for k = 2, provides a level of confidence of approximately 95%.
Tensile Tester / Universal Testing Machine
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 4 to <100 Newton | ± 0.06 Newton |
| 100 to <1,000 Newton | ± 0.47 Newton |
Standard / Technique / Method / EquipmentISO 7500-1:2004 (E) /Standard weight. & Standard Loadcell.
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 50 mm./min | ± 0.5 mm./min |
| 100 mm./min | ± 0.5 mm./min |
| 200 mm./min | ± 0.5 mm./min |
| 300 mm./min | ± 0.5 mm./min |
| 400 mm./min | ± 0.5 mm./min |
| 500 mm./min | ± 0.5 mm./min |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-17 Verification by Standard stopwatch and Vernier caliper.
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 100 % (Gauge 2 cm.) | ± 1.5 % |
| 200 %(Gauge 2 cm.) | ± 1.5 % |
| 300 %(Gauge 2 cm.) | ± 1.5 % |
| 400 %(Gauge 2 cm.) | ± 1.5 % |
| 500 %(Gauge 2 cm.) | ± 1.5 % |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-17 by direct comparison to standard gauge block.
Calibration and measurement capability is expressed as an uncertainty (±) which, for k = 2, provides a level of confidence of approximately 95%.
Melt Indexer
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 190 to 300 °C | ± 0.44 °C |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-21 by direct comparison to standard Thermometer.
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 190 to 350 °C | - °C |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-11 Verification by Standard stopwatch.
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 100.0 to 500.0 g. | ± 0.2 g. |
| 500.0 to 1000.0 g. | ± 0.3 g. |
| 1000.0 to 10000.0 g. | ± 0.3 g. |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-01 by comparison with standard weight.
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 8.890 mm. | ± 0.002 mm. |
| 9.474 mm. | ± 0.002 mm. |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-11 by comparison with standard Micrometer.
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 8.000mm. | ± 0.002 mm. |
| 9.500mm. | ± 0.002 mm. |
| 2.095mm. | ± 0.002 mm. |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-11 by comparison with standard Micrometer and comparison to standard Diameter gauge.
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 2.090mm. | ± 0.001 mm. |
| 2.100mm. | ± 0.001 mm. |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-11 by comparison with standard Micrometer.
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 9.550 mm. | ± 0.003 mm. |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-11 by comparison with standard Diameter gauge.
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 1.00-20.00 mm. | ± 0.008 mm. |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-12 by comparison with standard Micrometer head.
Calibration and measurement capability is expressed as an uncertainty (±) which, for k = 2, provides a level of confidence of approximately 95%.
HDT / Vicat Tester
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 50 to 100 °C | ± 0.27 °C |
| >100 to 200 °C | ± 0.35 °C |
| >200 to 350 °C | ± 0.45 °C |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-05 by direct comparison to standard Thermometer.
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 50 to 300 °C | 0.45 °C |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-14 Verification by standard Thermometer and Standard stopwatch.
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 100.0 to 500.0 g. | ± 0.2 g. |
| 500.0 to 1000.0 g. | ± 0.3 g. |
| 1000.0 to 10000.0 g. | ± 0.3 g. |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-01 by comparison with standard weight.
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| – 0.5 to 2.5 mm. | ± 0.003 mm. |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-13 by comparison with Micrometer Head.
Calibration and measurement capability is expressed as an uncertainty (±) which, for k = 2, provides a level of confidence of approximately 95%.
Impact Tester
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 1.0- <10.0 N.m | ± 0.1 % |
| 10.0- <20.0 N.m | ± 0.1 % |
| 20.0- <30.0 N.m | ± 0.1 % |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-10 by direct measurement to standard Electronic balance.
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| - All hammer | ± 0.7 % |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-10 by measurement to standard Stopwatch.
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 150 -600 mm. | ± 1.30 % |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-10 by measurement to standard Stopwatch.
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 20.0- 60.0 mm. | ± 0.01 %. |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-10 by measurement with standard Vernier caliper.
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 200.0- 610.0 mm. | ± 0.25 % |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-10 by measurement with standard Vernier caliper.
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 0.0- 150.0 degree. | ± 0.2 degree. |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-10 by comparison with standard Gauge block.
Calibration and measurement capability is expressed as an uncertainty (±) which, for k = 2, provides a level of confidence of approximately 95%.
Oven (Geer Oven)
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 100 – <150 °C | ± 1.50 °C |
| 150 – <200 °C | ± 1.80 °C |
| 200 – <250 °C | ± 2.10 °C |
| 250 – <300 °C | ± 2.50 °C |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-09 by direct comparison to standard Thermometer.
| Parameter / Range / Item | Calibration measurement capability |
|---|---|
| 0.3 to 10.0 m/s | ± 1.20 m/s |
Standard / Technique / Method / EquipmentIn-house method WI 7.2-09 by direct comparison to standard Anemometer.
Calibration and measurement capability is expressed as an uncertainty (±) which, for k = 2, provides a level of confidence of approximately 95%.
Ask us for the full accredited scope document.