- ISO 6502
- ASTM D5289
- DIN 53529
MDR 3000 Moving Die Rheometer
MonTech · Moving Die Rheometer
The MDR 3000 is the industry standard for measuring the viscoelastic properties of polymers and elastomeric compounds before, during and after cure. The acquired data is gives advanced information about processability, cure characteristics, cure speed, as well as the behavior of the compound after-cure at fixed, user selectable strain rates.
Datasheets
- MonTech General Cataloguemontech-product-catalogue-2023.pdfDownload PDF
- MonTech Specimen and Die cuttingmontech-specimen-and-die-cutting-2021.pdfDownload PDF
- MDR 3000 datasheetmoving-die-rheometer-2021.pdfDownload PDF
Specification
| International standards | ISO 6502, ASTM D 5289, DIN 53529, ... |
|---|---|
| Die configuration | Biconical, closed die system, sealed |
| Drive system | Direct, wearless servo drive system |
| Oscillation frequency | 1.667 Hz (100 cpm) |
| Oscillation strain | +/- 0.01° to 5° (+/- 0.14% to 70%), Programmable via Software Optional “M” version with mechanically fixed oscillation angle |
| Temperature range | Ambient to 232 °C |
| Measured Data | Torque (dNm, lbf.in, kgf.cm), Temperature (°C, °F), Pressure (bar, kg per cm²), Time (min - min / min - sec / sec), Shear rate (1/s, rad/s), Cure rate (1/min, 1/sec) |
| Calculated Data | S΄, S˝, S * , tan δ, phase angle, cure rate, ... |
| Die gap | 0.45 mm nominal |
| Sample volume | approx. 4.5 cm³ |
| Closing system | Soft closing to prevent film tearing and damage of test sample |
| Torque range | 0.01 to 250 dNm |
| Temperature control system | Ambient to 232 °C, precision +/- 0.03 °C, Max. heating rate: 85°C/min digital, microprocessor controlled |
| Temperature check system | Recordings of the temperature gradient on the screen, microprocessor monitored |
| Data Interface | Ethernet (10/100 MBit), USB (int.), CF card (int.), RS232 (opt.) |
| Data points | Over 3500 data points available for each test Including: S‘ Min, S‘ Max, TS 1, TS 2, TC 10, TC 30, TC 50, TC 90 |
| Pneumatics | min. 4.5 Bar / 60 psi |
| Electrical | 200-240 V, 6 Amps, 47-63 Hz, Single phase |
| Instrument options | - "M" version with mechanical drive system - Instrument control panel with 5" touchscreen display and printer - Adapted transducers for high or low-viscosity torque range - Normal force / Pressure measurement - Double channel forced air cooling system - Extended thermal range to +300°C - Autoloader 5 or 10 sample linear - Autoloader with 24, 48 or 100 sample tray or tray changers - R-VS 3000 constant volume sample cutter - Label Printer - Signal tower |
| Normal force / Pressure (opt.) | 0 to 10.000 kPa (1.450 lbin), Auto tare |
Model comparison
| MDR 3000 Basic | MDR 3000 | MDR 3000 Professional | D-MDR 3000 | D-RPA 3000 | |
|---|---|---|---|---|---|
| Instrument description | Compact, entry model instrument for repetitive quality control testing as well as simple R&D applications | High-end instrument for QC and static R&D testing for a broad range of materials (from LSR, all kinds of elastomers to composite materials) | Entry model instrument in the Rubber Process Analyzer (RPA) technology. The MDR 3000 Prof. can be operated either in static testing mode (MDR) or also in dynamic testing mode (RPA). | Top model for advanced dynamic testing of all kinds of rubber, rubber-like, curing or crosslinking materials. This includes TPE, TPV, LSR, Composite materials, Polyolefins, ... | High-end instrument for demanding static and dynamic testing in QC and R&D with a huge set of customizing options. This enables testing capabilities that have not been possible to measure in a Rheometer before, such as die gap regulation, High-speed data acquisition, FT-Rheology, low-temperature cooling, … |
| International standards | ISO 6502, ASTM D 5289, DIN 53529 | ISO 6502, ASTM D 5289, DIN 53529 | ISO 13145, ISO 6502, ASTM D 5289, ASTM D 6204, ASTM D 6601, ASTM D 6048, ASTM D 7050, ASTM D 7605, ASTM D 8059, DIN 53529 | ISO 13145, ISO 6502, ASTM D 5289, ASTM D 6204, ASTM D 6601, ASTM D 6048, ASTM D 7050, ASTM D 7605, ASTM D 8059, DIN 53529 | ISO 13145, ISO 6502, ASTM D 5289, ASTM D 6204, ASTM D 6601, ASTM D 6048, ASTM D 7050, ASTM D 7605, ASTM D 8059, DIN 53529 |
| Die configuration | Biconical, closed die system, sealed | Biconical, closed die system, sealed | Biconical, closed die system, sealed | Biconical, closed die system, sealed | Biconical, closed die system, sealed |
| Drive system | Mechanical, brushless DC eccentric drive | Direct, wearless servo drive system | Direct, wearless servo drive system with ceramic bearings | Direct, wearless servo drive system with ceramic bearings | Advanced wearless servo drive system with ceramic bearings |
| Oscillation frequency | 1.667 Hz | 1.667 Hz | 0.001 Hz to 33 Hz (0.001 Hz to 50 Hz optional) | 0.001 Hz to 100 Hz | 0.001 Hz to 100 Hz |
| Oscillation strain | +/- 0.1°, 0.2°, 0.5°, 1°, 3° or 7.17° Mechanically adjustable | +/- 0.01° to 5° Programmable via Software | +/- 0.01° to 20° (+/- 0.01° to 90° optional) | +/- 0.001° to 180° | +/- 0.0001° to 360° |
| Temperature range | Ambient to 232 °C | Ambient to 232 °C | Ambient to 232 °C | Ambient to 232 °C | Ambient to 232 °C |
| Measured Data | Torque, temperature, frequency / Optional: Normal force | Torque, temperature, frequency, strain / Optional: Normal force | Torque, temperature, frequency, strain / Optional: Normal force | Torque, temperature, frequency, strain / Optional: Normal force | Torque, temperature, frequency, strain / Optional: Normal force, die gap, die pressure |
| Calculated Data | S΄, S˝, S *, tan δ | S΄, S˝, S *, tan δ | S΄, S˝, S *, G΄, G˝, G *, tan δ, η΄, η˝ and η* | S΄, S˝, S *, G΄, G˝, G *, tan δ, η΄, η˝ and η* | S΄, S˝, S *, G΄, G˝, G *, tan δ, η΄, η˝ and η* |
Scroll the table sideways for every column