TH
  • ISO 812
  • ISO 974
  • ASTM D746

Brittleness Temperature Tester (Brittle Point Temperature Tester) | 525

ModelNo. 525

Toyo Seiki · Heat Resistance / Low Temperature / Heat Aging

The Brittleness Temperature Tester evaluates specimens immersed in liquid at low temperature. Impact is applied by hammer at constant speed and evaluation is made from generation of cracks. Specimen Clamps OPTIONS

Datasheets

Specification

ISO 812Rubber, vulcanized or thermoplastic — Determination of low-temperature brittleness
ISO 974Plastics — Determination of the brittleness temperature by impact
ASTM D746Standard Test Method for Brittleness Temperature of Plastics and Elastomers by Impact

Model comparison

ISO 812, ISO 974, ASTM D746
ModelFSS
Cooling methodRefrigeratorDry ice
Heat transfer mediumEthanol, approx. 10L
Number of clamping specimens5 specimens per specimen clamp
Temperature range-70°C to +40°C
Power supplySingle-phase, AC100V, 50/60Hz, 3kVA (or specify) / (AC230V, 115V etc. are available upon request using optional transformer)Single-phase, AC100V, 50/60Hz, 1.5kVA (or specify) / (AC230V, 115V etc. are available upon request using optional transformer)
DimensionsW650 × D700 × H1250mmW580 × D710 × H470mm
Weight (approx.)190kg110kg

Scroll the table sideways for every column

Specimen Clamps
CodeTest standardsNumber of specimensClamping type
CLAMP 3• ASTM D746 (Specimen type I) / • ASTM D746 (Specimen type II)5Individual clamping
CLAMP 4• ISO 974 / • ASTM D746 (Specimen type III)10Group clamping
CLAMP 5• ISO 812 (Type B test piece)5Individual clamping

Scroll the table sideways for every column

OPTIONS
For model FSFor model S
Hose for ethanol dischargeHS-FSHS-S
Digital thermometer—–D-THRM
Torque screwdriverTQDRV

Scroll the table sideways for every column