- ISO 11443
- ASTM D3835
Capillary Rheometer | Capilograph™ | 583
Toyo Seiki · Melt Flow Index (MFR, MVR) / Melt Viscosity
The Capilograph measures melt viscosity of polymers by detecting the shear rate and shear stress of melt polymer when it flows out of the capillary die. Die swell and melt strength can also be measured by equipping it with optional attachments. The twin bore barrel model (F-2) allows easy Bagley & Weissenberg-Rabinowitsch correction (for determination of true viscosity) or elongational viscosity calculation.
Datasheets
Specification
| ISO 11443 | Plastics — Determination of the fluidity of plastics using capillary and slit-die rheometers |
|---|---|
| ASTM D3835 | Standard Test Method for Determination of Properties of Polymeric Materials by Means of a Capillary Rheometer |
Model comparison
| Item | Value |
|---|---|
| ISO 11443 | Plastics — Determination of the fluidity of plastics using capillary and slit-die rheometers |
| ASTM D3835 | Standard Test Method for Determination of Properties of Polymeric Materials by Means of a Capillary Rheometer |
Scroll the table sideways for every column
| Model | F-1 | F-2 |
|---|---|---|
| Temperature range | 60 to 400°C (500°C optional) | |
| Temperature accuracy | ±1°C (Less than 300°C) / Within ±2°C (300°C or more) | |
| Number of bores | Single | Twin |
| Bore diameter | Selection: / • ∅9.55mm / • ∅15mm | ∅15mm |
| Barrel length | Length: 350mm, effective length: 250mm | |
| Piston speed | 0.1 to 1500mm/min. | |
| Force range | • 25kN max (bore dia. ∅9.55mm) / • 50kN max (bore dia. ∅15mm) | 25kN max (bore dia. ∅15mm) |
| Power supply | Three-phase, AC200V, 50/60Hz, 5.2kVA (Main unit) | |
| Compressed air requirement | 0.5MPa | |
| Dimensions | W1060 × D1100 × H1950 to 2360mm | |
| Weight | Approx. 650kg |
Scroll the table sideways for every column